SKU:

Three Gas Analyzer (F-950)

Measures ethylene, CO2 and O2. 0-200ppm C2H4 measurement range.

Measures ethylene, CO2 and O2. 0-200ppm C2H4 measurement range.

Loading...

Product Details

The F-950 Three Gas Analyzer measures levels of Ethylene, CO2 and O2 to maintain optimal produce quality at every phase of the supply chain. The F-950’s versatile design allows it to be be scaled to many environments, from cold storage to transportation container.

 

Features

  • Measures C2H4, CO2 and O2
  • Displays results in under 30 seconds
  • Past records stored on the device
  • Handheld and lightweight
  • Internal data logging and storage (4GB SD card)
  • Battery lasts for 8+ hours
  • Display easily viewed in dimly lit environments
  • PolarCept!TM water filter for removing non-ethylene hydrocarbons
  • The F-950 supports English, Spanish, and Portuguese on the instrument display

Specification

Sampling Flow Rate 70mL/min
Total Sampling Time 80 seconds
Data Saving Automated, 1 second intervals
Display Sunlight visible transflective LCD
Operating Environment 0° C – 50° C (0-90 % humidity non-condensing)
Operating Atmospheric Pressure 500-1200mbar
Dimensions 18 cm x 13.5 cm x 5.5 cm
Weight 0.95 kg
Enclosure Powder coated aluminum
Battery Life 8.5 hours
Power Source Removable rechargeable lithium-ion battery
PC Interface USB and WI-FI SD card + Bluetooth
Data Recorded with Each Measurement Ethylene, CO2 and O2 concentrations, date, time, RH, GPS location

 

Trigger Mode Measurements

C2H4 CO2 O2
Sensor Type Electrochemical Infrared Sensor, Pyroelectric detector Electrochemical
Range 0-200ppm 0-100% 0-100%
Resolution 0.1ppm 0.01% 0.10%
Lower Detection Limit 0.15ppm 0.50% 0.30%
Sampling Volume 35 ml 12 ml 12 ml
Response Time 35 seconds 12 seconds 12 seconds
Accuracy (absolute) 0.15 ppm 0.50% 0.30%
Accuracy (relative) 5% 3% 2%
Offset Calibration Weekly Weekly Weekly
Span Calibration 3 months 12 months 6 months
Sensor Replaceable (by user) yes no yes
Lifetime 1 year 5 years 2 years

 

Continuous Mode Measurements

C2H4 CO2 O2
Sensor Type Electrochemical Infrared Sensor, Pyroelectric detector Electrochemical
Range 0-200ppm 0-100% 0-100%
Resolution 0.1ppm 0.01% 0.10%
Lower Detection Limit 0.2ppm 0.01% 0.10%
Response Time 80 seconds 6 seconds 12 seconds
Accuracy (absolute) 0.2 ppm 0.01% 0.10%
Accuracy (relative) 5% 3% 2%
Offset Calibration Weekly Weekly Weekly
Span Calibration 3 months 12 months 6 months
Sensor Replaceable (by user) yes no yes
Lifetime 1 year 5 years 2 year

 

 

Visit www.felixinstruments.com to learn more about the F-950.

Video