Ripen It! Gas Analyzer (F-960)

Measures ethylene, CO2, and O2. 0-1000ppm C2H4 measurement range

Measures ethylene, CO2, and O2. 0-1000ppm C2H4 measurement range

Loading...

Product Details

The F-960 Ripen It! Gas Analyzer rapidly measures ethylene, CO2, and O2 to assist with decision making throughout the produce ripening process.

 

Features

  • Measures C2H4 , CO2 and O2
  • Displays results in under 30 seconds
  • Past records stored on the device
  • Handheld and lightweight
  • Battery lasts for 8+ hours
  • Wi-Fi capabilities
  • Display easily viewed in dimly lit environments

Specification

Specifications

Sampling Flow Rate 70mL/min
Total Sampling Time 90 seconds
Data Saving Automated, 1 second intervals
Display Sunlight visible transflective LCD
Operating Environment 0º C - 50º C (0-90% humidity non-condensing)
Operating Atmospheric Pressure 500-1200mbar
Dimensions 18 cm x 13.5 cm x 5.5 cm
Weight 0.95 kg
Enclosure Powder coated aluminium
Battery Life 8.5 hours
Power Source Removable rechargeable lithium-ion battery
PC Interface USB and WI-FI SD card + Bluetooth
Data Recorded with Each Measurement Ethylene, CO2 and O2 concentrations, date, time, RH, GPS location

 

Trigger Mode Measurements

  C2H4 CO2 O2
Sensor Type Electrochemical Infrared Sensor, Pyroelectric detector Electrochemical
Range 0-1000 ppm 0-100% 0-100%
Resolution 1 ppm 0.01% 0.10%
Lower Detection Limit 10 ppm 0.50% 0.30%
Sampling Volume 35 ml 12 ml 12 ml
Response Time 35 seconds 12 seconds 12 seconds
Accuracy (absolute) 10 ppm 0.50% 0.30%
Accuracy (relative) 5% 2% 3%
Offset Calibration Weekly Weekly Weekly
Span Calibration 3 months 12 months 6 months
Sensor Replaceable (by user) yes no yes
Lifetime 1 year 5 years 2 years

 

Continuous Mode Measurements

  C2H4 CO2 O2
Sensor Type Electrochemical Infrared Sensor, Pyroelectric detector Electrochemical
Range 0-1000 ppm 0-100% 0-100%
Resolution 1 ppm 0.01% 0.10%
Lower Detection Limit 2 ppm 0.01% 0.10%
Response Time 90 seconds 6 seconds 12 seconds
Accuracy (absolute) 2 ppm 0.01% 0.10%
Accuracy (relative) 5% 3% 2%
Offset Calibration Weekly Weekly Weekly
Span Calibration 3 months 12 months 6 months
Sensor Replaceable (by user) yes no yes
Lifetime 1 year 5 years 2 years

 

Visit www.felixinstruments.com to learn more about the F-960 Ripen It! Gas Analyzer.