SKU:

Ozone Monitor (Model 106-L)

Model 106-L measuring range from low ppb levels up to 100 ppm (100,000 ppb). This highly accurate ozone monitor (1.5 ppb) has a low limit of detection (3.0 ppb for 10 second measurements) and is certified by the EPA as a FEM for monitoring for compliance with the US Clean Air Act:  EQOA-0914-218.

Model 106-L measuring range from low ppb levels up to 100 ppm (100,000 ppb). This highly accurate ozone monitor (1.5 ppb) has a low limit of detection (3.0 ppb for 10 second measurements) and is certified by the EPA as a FEM for monitoring for compliance with the US Clean Air Act:  EQOA-0914-218.

Loading...

Description

Measurement Principle UV Absorption at 254 nm, single beam
Measurement Interval 2 s
Linear Dynamic Range 0-100,000 ppb (0-100 ppm)
Resolution 0.1 ppb
Precision (1σ for 10-s average; aka rms noise) Greater of 1.5 ppb or 2% of Reading
Limit of Detection (10-s average, 2σ) 3.0 ppb
Accuracy Greater of 1.5 ppb or 2% of Reading
Baseline Drift < 3 ppb/day, < 6 ppb/year
Sensitivity Drift < 1%/day, < 3%/year
Calibration NIST Traceable; Annual calibration recommended
Measurement Time, Frequency 2 s, 0.5 Hz
Data Averaging Options 10 s, 1 min, 5 min, 1 hr
Response Time, 100% of Step Change For 2-s output: 4 s, 2 data points
For 10-s output: 20 s, 2 data points
Adaptive Filter Available, user-specified parameters
Data Logger Capacity 32,736 lines (10 s avg. = 3.8 days; 5 min avg = 113 days)
Data Transfer Baud Rate 2400, 4800, 19200
Ozone Units ppb, pphm, ppm, µg m-3, mg m-3
Temperature Units oC, K
Pressure Units mbar, torr
T and P Corrected Yes
Operating Temperature Range 0 to 50oC
Operating Altitude Range 0-13.5 km (0-25 km option)
Flow Rate Minimum required: 0.6 L/min; Nominal: 1 L/min; Maximum: 1.5 L/min
Digital Data Outputs USB, RS232, LCD Display
Analog Data Outputs 0-2.5 V Analog, 4-20 mA, User-Scalable in Menu
Power Requirements 100-240 VAC, 50/60 Hz
11-28 V DC, nominally 500 mA at 12 V DC, 6 watt
Relays with 2 Setpoints Two available:
Relay 1 responds based on user’s ozone set points
Relay 2 responds based on user’s ozone set points OR responds based on diagnostics (T, P, flow, lamp voltage)
Four relays for Industrial models and in 3-channel and 6-channel optional configurations
Bluetooth Option Available Yes
Flow-Through Option Available Yes
Multi-Channel Options Available Yes, 3-channel and 6-channel configurations
Size Standard Case: 3.6 x 7.9 x 9.4 inches (9 x 20 x 24 cm)
OEM Version: 2.5 x 7 x 9 in (6.2 x 17.8 x 22.9 cm)
Industrial Version: 16 x 14 x 9.3 in (40.7 x 35.7 x 23.6 cm) H x W x D
Weight Standard Case: 4.2 lb (1.9 kg)
OEM Version: 2.8 lb (1.3 kg)
Industrial Version: ~14 lb (6.3 kg) depending on multi-channel options selected
Options Battery, Particle Filter, Bluetooth, Exhaust Port, Flow-Through Configuration, Multi-Channel Configurations, High-Altitude Package

Product Details

  • Designed for industrial ozone applications to cover four different ozone concentration ranges by varying the optical path length. 
  • Highly robust and easy to service; all components are mounted directly to the printed circuit board with few wire connection
  • Available in the three following configurations: Standard Enclosure, Industrial Enclosure, or with no enclosure (OEM version)
  • Industrial/NEMA Enclosure is available with single channel, three channel, or six channel sampling manifolds (106-H only available as single channel)

 

106 Series Comparison:

Model Range Resolution Precision & Accuracy
106-L 0-100 ppm 0.0001 ppm (0.1 ppb) Higher of 0.0015 ppm (1.5 ppb) or 2% of reading
106-M 0-1,000 ppm 0.01 ppm Higher of 0.01 ppm or 2% of reading
106-MH 0-10,000 ppm (0-1 vol%) 0.1 ppm Higher of 0.05 ppm or 2% of reading
106-H 0-20 wt% O2 or Air
0-14 vol%
0.01 wt%
0.01 vol%
Higher of 0.01 wt% or 2% of reading